Diagnostic Accuracy of Contrast-Enhanced 16-Slice Multidetector CT for Detecting Mediastinal Invasion in Esophageal Cancer: A Prospective Cross-Sectional Study
DOI:
https://doi.org/10.37185/Keywords:
Diagnostic Imaging, Esophageal Neoplasms, Mediastinum, Multidetector Computed Tomography, Neoplasm Invasiveness, Sensitivity and SpecificityAbstract
Objective: To determine the diagnostic performance of contrast-enhanced 16-slice multidetector computed tomography (MDCT) for detecting mediastinal invasion in esophageal carcinoma, using histopathology as the reference standard.
Study Design: Prospective cross-sectional diagnostic-accuracy study.
Place and Duration of Study: This study was conducted at the Department of Radiology, Memon Medical Institute Hospital, Karachi, Pakistan, from 15th February 2024 to 15th July 2024.
Methods: Consecutive patients aged 20–80 years with biopsy-proven esophageal carcinoma were enrolled. Contrast-enhanced MDCT of the neck, chest, and upper abdomen was performed on a Toshiba Aquilion 16 scanner. A radiologist blinded to the reference-standard findings classified mediastinal invasion as present or absent based on loss of fat planes or direct extension into adjacent mediastinal structures. Surgical resection provided histopathological confirmation. We calculated diagnostic estimates from a 2 × 2 table with exact 95% confidence intervals (CIs). Subgroup receiver-operating-characteristic (ROC) plots had one nontrivial operating point; we report the area under the curve (AUC) as balanced accuracy rather than as a threshold-optimization analysis.
Results: Histopathology confirmed mediastinal invasion in 189 of 271 patients (69.7%). MDCT yielded 160 true positives, 65 true negatives, 17 false positives, and 29 false negatives. Sensitivity was 84.7% (95% CI, 78.7%–89.5%), specificity 79.3% (68.9%–87.4%), positive predictive value 90.4% (85.1%–94.3%), negative predictive value 69.1% (58.8%–78.3%), and overall accuracy 83.0% (78.0%–87.3%). Exploratory binary-classification AUCs ranged from 0.779 to 0.863 across sex and tumor-size subgroups. Although eligibility extended to 80 years, the enrolled cohort was 20–50 years old (mean, 35.64 ± 7.99 years).
Conclusion: Contrast-enhanced 16-slice MDCT showed good sensitivity, specificity, and positive predictive value for mediastinal invasion. Its lower negative predictive value means that a negative MDCT result should not be used alone to exclude microscopic invasion when clinical suspicion remains.
How to cite this: Azmatullah U, Khuwaja MR, Irshad R, Abbas S, Soomro MI. Diagnostic Accuracy of Contrast-Enhanced 16-Slice
Multidetector CT for Detecting Mediastinal Invasion in Esophageal Cancer: A Prospective Cross-Sectional Study. Life and Science. 2026;
7(4): 472-478. doi: http://doi.org/10.37185/LnS.1.1.987